The Shodex RI-500 series is a versatile and high sensitive RI detectors that can be used with various manufactures 'HPLC system'. It consists of a color-LCD, an automatic start-up function, and validation wizard. "The detector series are available as Analytic, semi-micro and preparative line up. (Features)
· The automatic start-up function controls the complicated operations such asblank substitution and baseline atabilization automatically.
· The validation wizard enables easy component validation.
· Improved double temparature control system shortens the required warm-up time and provides stable background.
· The leak sensor automatically stops the pump in case of solvent leakage.
· External input and output terminals and LAN (RJ45) communication ports can be used to connect various HPLC systems for an advanced automation.
· Usable in UHPLC systems.
(Specifications)
Model | Analysis | Semi-preparative | Micro | |
---|---|---|---|---|
RI-501 | RI-502 | RI-504 | ||
Measuremennt method |
Deflection type
|
|||
Flow cell |
2 chanber-type
|
|||
Reflactive index range |
1.00 ~ 1.75
|
|||
Linearity range |
600μRIU
|
6000μRIU
|
600μRIU
|
|
Recorder output range selection |
0.25 ~ 512µRIU
|
2.5 ~ 5120µRIU
|
0.25 ~ 512µRIU
|
|
Integrator output range selection |
128/512µRIU/FS
|
1280/5120µRIU/FS
|
128/512µRIU/FS
|
|
Noise |
2.5nRIU
(Pure water, response : 1.5sec) |
25nRIU
(Pure water, response : 1.5sec) |
5nRIU
(Pure water, response : 1.5sec) |
|
Drift |
0.2μRIU/H
|
2μRIU/H
|
0.2μRIU/H
|
|
Response |
0.1 , 0.25 , 0.5 , 1 , 1.5 , 2 , 3 , 6sec
|
|||
Auto zero |
Full auto zero
|
|||
Auto zero range |
All range
|
|||
Integrator output |
0 ~ 1V
|
|||
Recorder output |
0 ~ 10mV/FS
|
|||
External signal input |
Zero / Purge
|
|||
External signal output |
READY / LEAK / ERROR
|
|||
Flow cell volume |
8μL
|
8μL
|
2.5μL
|
|
Temperature control system |
OFF, 30~55°C (double Temperature control)
|
|||
Usable flow rate |
0.2 ~ 3mL/min
|
1 ~ 50mL/min
|
0.2 ~ 1mL/min
|
|
Maximum back pressure |
50kPa
|
|||
Dead volume |
IN to Cell : 60μL
Cell to OUT : 630μL |
IN to Cell : 120μL
Cell to OUT : 540μL |
IN to Cell : 10μL
Cell to OUT : 385μL |
|
Communication port |
LAN
|
|||
Power source |
AC100 ~ 240V±10%, 50/60Hz, 150VA
|
|||
Dimensions and weight |
260W × 420D × 165H (mm) 12.5kg
|
|||
CE Marking conformity | obtained | |||
RoHS conformity | Compliance |
Please select your location to continue:
For all other locations please click “close” and return to the homepage.
This website stores cookies on your computer. These cookies are used to collect information about how you interact with our website and allow us to remember you. We use this information in order to improve our services. If you are happy to accept these cookies please click “Accept” or simply continue to use our website. To find out more about the cookies we use, see “More Information”.