Solvent Compatibility of EV-2000 (1)

The in-column solvent of CLNpak EV-2000 can be replaced easily without affecting column performance.

The column shipping solvent of EV-2000 column is ethyl acetate/cyclohexane(3/7).
(1) After pumping acetone/cyclohexane (2/8) at 2 mL/min for 2 h.
(2) After pumping ethyl acetate/cyclohexane (2/8) at 2 mL/min for 2 h.
(3) After pumping acetone/cyclohexane (2/8) at 2 mL/min for 2 h.
(4) After pumping ethyl acetate/cyclohexane (2/8) at 2 mL/min for 2 h.


Sample :
1. Corn oil
2. Fluvalinate
3. Chinomethionate


Columns      : Shodex CLNpak EV-G (20 mm I.D. x 100 mm) + EV-2000 (20 mm I.D. x 300 mm) 
Flow rate    : 4.0 mL/min
Detector     : UV (254 nm)
Column temp. : 25 °C

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications

Applications (Related Information)