Organic SEC (GPC) Column: Ultra-rapid Analysis

Scroll
HK-400
  • Newly developed styrene divinylbenzene copolymer monodisperse particles
  • Analysis time is reduced to about a sixth of conventional column's analysis time
  • Low column pressure even under high flow rate does not require its use with UHPLC device
  • The amount of solvent used is reduced to about a sixth
  • Fulfills USP L21 requirements
(High performance semi-micro columns)   [Housing material: SUS]
HK-400 series is recommended to be used with semi-micro type devices.Scroll
Product code Product name Plate number
(TPN/column)
Particle size (μm) Column size (mm)
I.D. x Length
F6025010 GPC HK-401   ≥ 9,000 3 4.6 x 150
F6025020 GPC HK-402 ≥ 12,000 3 4.6 x 150
F6025030 GPC HK-403   ≥ 9,000    3.5 4.6 x 150
F6026040 GPC HK-404L   ≥ 9,000    3.5 4.6 x 150
F6025050 GPC HK-405   ≥ 7,000 3 4.6 x 150
F6025060 GPC HK-406   ≥ 5,000    6.5 4.6 x 150
HK-404L is a mixed-gel column capable of analyzing samples over a wide range of molecular weight distribution.
Scroll
Product name Maximum pressure
(MPa/Column)
Flow rate (mL/min) Temperature (℃) Shipping solvent
Recommended Maximum Recommended Maximum
GPC HK-401 25 0.3 - 1.0 2.0 40 60 Tetrahydrofuran (THF)
GPC HK-402 20 0.3 - 1.0 1.5 40 60 Tetrahydrofuran (THF)
GPC HK-403 25 0.3 - 1.0 2.0 40 60 Tetrahydrofuran (THF)
GPC HK-404L 25 0.3 - 1.0 2.0 40 60 Tetrahydrofuran (THF)
GPC HK-405 25 0.3 - 1.0 2.0 40 60 Tetrahydrofuran (THF)
GPC HK-406 25 0.3 - 1.0 2.0 40 60 Tetrahydrofuran (THF)
(Guard filter for GPC HK series)Scroll
Product code Product name Contents
F6700200 GPC HK-G One holder and one filter
F6700100 GPC HK-G filter 3 filters
 
- Removes sample-origin insoluble components.
- Attach directly to the analytical column.
Usable solventsScroll
Solvents Product name
HK-401 HK-402 HK-403 HK-404L HK-405 HK-406
Chloroform
N,N-Dimethylformamide (DMF)
Toluene
Hexafluoroisopropanol (HFIP) ×
30 % HFIP/Chloroform
✓: Can be used ×: Cannot be used
Target molecular weight range and exclusion limitPlease use the following tables for reference purposes only when selecting columns. Scroll
Product name Measured with polystyrene (eluent: THF)
Target molecular weight range Exclusion limit
GPC HK-401 100 - 1,500 2,000
GPC HK-402 200 - 10,000 20,000
GPC HK-403 2,000 - 70,000 100,000
GPC HK-404L 100 - 1,000,000 1,000,000
GPC HK-405 10,000 - 2,500,000 4,000,000
GPC HK-406 30,000 - 8,000,000 10,000,000

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Related Products

Product Information (Related Information)