The in-column solvent of CLNpak EV-2000 can be replaced easily without affecting column performance.
The column shipping solvent of EV-2000 column is ethyl acetate/cyclohexane(3/7).
(1) After pumping acetone/cyclohexane (2/8) at 2 mL/min for 2 h.
(2) After pumping ethyl acetate/cyclohexane (2/8) at 2 mL/min for 2 h.
(3) After pumping acetone/cyclohexane (2/8) at 2 mL/min for 2 h.
(4) After pumping ethyl acetate/cyclohexane (2/8) at 2 mL/min for 2 h.
Sample :
1. Corn oil
2. Fluvalinate
3. Chinomethionate
Columns : Shodex CLNpak EV-G (20 mm I.D. x 100 mm) + EV-2000 (20 mm I.D. x 300 mm) Flow rate : 4.0 mL/min Detector : UV (254 nm) Column temp. : 25 °C