Durability Against Alkaline Solvents (UB Series)

Alkali resistance was tested by passing 1 M NaOH aqueous solution for 24 hours. There was no variation in distribution coefficients of various proteins, including acidic, basic, and hydrophobic proteins, before and after passing NaOH aqueous solution, confirmed UB series column’s high alkaline durability.

Alkali resistance test conditions
Column
Shodex UB-100 (8.0 mm I.D. x 300 mm)
Eluent
1M NaOH aq.
Flow rate
0.25 mL/min
Column temp.
25 ℃
Passing time
24 h

Sample:

  1. Lysozyme
  2. Aprotinin
  3. BSA
  4. γ-Globulin
  5. Thyroglobulin
  6. IgM
  7. Hemocyanin
Alkali Resistance of UB series
Column
Shodex UB-100 (8.0 mm I.D. x 300 mm)
Eluent
50 mM Sodium phosphate buffer (pH 7.0) + 0.3 M NaCl
Flow rate
0.6 mL/min
Detector
UV (280 nm)
Column temp.
25 ℃