Silica-based Aqueous SEC (GFC) Columns: PROTEIN KW-800

Scroll
KW-800
  • Silica-based packed columns for aqueous SEC (GFC) analysis
  • Suitable for the analysis of proteins and enzymes
  • Fulfills USP-NF L20, L33, and L59 requirements
(Standard columns)   [Housing material: SUS]Scroll
Product code Product name *Plate number
(TPN/column)
Particle size (μm) Column size (mm)
I.D. x Length
F6989000 PROTEIN KW-802.5 ≥ 21,000 5
8.0 x 300
F6989103 PROTEIN KW-803 ≥ 21,000 5 8.0 x 300
F6989104 PROTEIN KW-804 ≥ 16,000 7
8.0 x 300
F6700131 PROTEIN KW-G 6B (guard column) 7
6.0 x   50
* Measured with ethylene glycol
Scroll
Product name Maximum pressure
(MPa/Column)
Flow rate (mL/min) Temperature range
(℃)
pH range Shipping solvent
Recommended Maximum
PROTEIN KW-802.5 5.0 1.0 1.5 45 3 - 7.5 H2O
PROTEIN KW-803 5.0 1.0 1.5 45 3 - 7.5 H2O
PROTEIN KW-804 5.0 1.0 1.5 45 3 - 7.5 H2O
PROTEIN KW-G 6B - - - 45 3 - 7.5 H2O
Target molecular weight range and exclusion limitPlease use the following tables for reference purposes only when selecting columns. Scroll
Product name Measured with protein (eluent: phosphate buffer) Measured with pullulan (eluent: ultrapure water)
Target molecular weight range Exclusion limit Target molecular weight range Exclusion limit
PROTEIN KW-802.5 5,000 - 100,000 150,000   2,000 - 50,000 60,000
PROTEIN KW-803 10,000 - 700,000 *(1,000,000) 5,000 - 100,000 170,000
PROTEIN KW-804 30,000 - *(4,000,000) *(4,000,000) 20,000 - 300,000 500,000
*( ) Estimated value

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Related Products

Product Information (Related Information)

General Information on Packed Columns

introduction of column cleaning procedure and column test conditions etc.