Features of VG-50 (1) (Recovery of Reducing Sugar)

In general recovery rate of reducing sugars such as mannose, arabinose, and xylose using amino columns tend to be low. This is caused by the formation of Schiff base which makes reducing sugars to be adsorbed to the amino group on the packing material. However, HILICpak VG-50 is an amino column that provides high recovery of reducing sugars. This allows high sensitivity analysis of reducing sugars.

Sample: 5 µL
5 mg/mL each

  1. 1.Fructose
  2. 2.Mannose
  3. 3.Glucose
  4. 4.Sucrose
Column
Shodex HILICpak VG-50 4D, Slica-based amino columns from other manufacturers (4.6 mm I.D. x 150 mm each)
Eluent
H2O/CH3CN=20/80
Flow rate
0.6 mL/min (VG-50 4D), 1.0 mL/min (Silica-based amino column)
Detector
RI
Column temp.
40 ℃