Simultaneous Analysis of Anionic and Non-ionic Surfactants
OHpak
SB-802.5 HQ, an aqueous solvent SEC (GFC) column, was used to analyze an anionic surfactant (sodium dodecylbenzenesulfonate) and non-ionic surfactant (triton X-100) simultaneously. Owing to the ion exclusion mode, anionic surfactant elutes at first and it is separated from non-ionic surfactant.

Sample : 100μL
(A)
Sodium dodecylbenzenesulfonate(Soft type)
50mg/L
(B)
Triton X-100 (
Octylphenol ethoxylate) 1000mg/L
Column : Shodex OHpak SB-802.5 HQ (8.0mmID*300mm)
Eluent : H2O/CH3CN=70/30
Flow rate : 1.0mL/min
Detector : UV(254nm)
Column temp. : 60°C