Carbonate and Other Anions (2) (NI-424)

IC NI-424 has high theoretical plate number (more than 5,000). Carbonate and other anions can be separated well with NI-424.


Sample : 50micro-L
1. HCO3- 40.0mg/L
2. H2PO4- 4.0mg/L
3. F- 0.3mg/L
4. Cl- 0.5mg/L
5. NO2-1.2mg/L
6. Br- 2.0mg/L
7. NO3- 2.0mg/L
8. SO42- 2.0mg/L

Column       : Shodex IC NI-424 (4.6mmID*100mm)
Eluent       : *18mM 4-Hydroxybenzoic acid + 2.8mM Bis-Tris + 2mM Phenylboronic acid + 0.005mM *2CyDTA(pH4.2)
Flow rate    : 1.0mL/min
Detector     : Shodex CD
Column temp. : 40deg-C
*1: The prescribed pH can be obtained by mixing the reagents in the composition described here.