Anion Standards (15) (Comparison of NI-424 and I-524A)

The seven main anions can be separated using a newly developed column, IC NI-424 quickly. The theoretical plate number of the column is more than double that of the existing column IC I-524A.
Comparison with I-524A column:
1) NI-424 column has achieved separation of phosphoric acid ion and fluoride ion which was difficult to sepatate using I-524A.
2) With NI-424, each peak is sharper, and separation of ions and peak balance have been improved. In particular, improvement has been made in the separation of chloride and nitrite ions making NI-424 columns suited to the analysis of a trace amount of nitrite ion contained in high-concentration chloride ion; demand for such analysis is high in the area of environmental analysis.


Sample :
1. H2PO4-
2. F-
3. Cl-
4. NO2-
5. Br-
6. NO3-
7. SO42-

(Left)
Column       : Shodex IC I-524A (4.6mmID*100mm)
Eluent       : *12.5mM Phthalic acid + 2.3mM Tris(pH4.2)
Flow rate    : 1.2mL/min
Detector     : Shodex CD
Column temp. : 40deg-C
(Right)
Column       : Shodex IC NI-424 (4.6mmID*100mm)
Eluent       : *18mM 4-Hydroxybenzoic acid + 2.8mM Bis-Tris + 2mM Phenylboronic acid + 0.005mM *2CyDTA(pH4.2)
Flow rate    : 1.0mL/min
Detector     : Shodex CD
Column temp. : 40deg-C
*1 The prescribed pH can be obtained by mixing the reagents in the composition described here.
*2 CyDTA: trans-1,2-Diaminocyclohexane-N,N,N',N'-tetraacetic acid