Influence of Temperature on Retention Time (SI-52 4E)
Please use this column with suppressor type ion chromatography system.
In analysis using IC
SI-52 4E, the retention times of some anions are very sensitive to
the column temperature. An example is showed in the figure below.
Bromide (Br
-) and chlorate (ClO
3-) are well separated at 45 deg-C or over,
while the separation of chlorite (ClO
2-) and bromate (BrO
3-) are
improved with lowering temperature.

Sample : 100micro-L
1.
F- (2mg/L)
2.
ClO2- (1mg/L)
3.
BrO3- (2mg/L)
4.
Cl- (10mg/L)
5.
NO2- (5mg/L)
6.
Br- (10mg/L)
7.
ClO3- (1mg/L)
8.
NO3- (30mg/L)
9.
PO43- (15mg/L)
10.
SO42- (40mg/L)
Column : Shodex IC SI-52 4E (4.0mmID*250mm)
Eluent : 3.6mM Na2CO3 aq.
Flow rate : 0.8mL/min
Detector : Suppressed conductivity