Hydrophobic Anion (SI-90 4E)
Please use this column with suppressor type
ion chromatography system.
Hydrophobic anions were analyzed using IC
SI-90 4E (a column
for anions analysis). If the acetone is added to eluent, hydrophobic anions are be eluted early. Moreover,
the separation of SO
32- and SO
42- improves.

Sample : 20micro-L
F-
Cl-
NO2-
Br-
NO3-
SO32-
PO43-
SO42-
S2O32-
SCN-
Column : Shodex IC SI-90 4E (4.0mmID*250mm)
Eluent : 1mM Na2CO3 + 4mM NaHCO3 + 5% Acetone
Flow rate : 1.5mL/min
Detector : Suppressed conductivity
Column temp. : Room temp.