Sodium Azide (1) (I-524A)
Sodium azide was analyzed using IC
I-524A (a column for anions analysis).
(Preparation of eluent(pH4.0))
(1) Put 0.415g of reagent grade phthalic acid and 0.278g of reagent grade tris(hydroxymethyl)aminomethane
into a 1 liter messflask.
(2) Dilute with purified water to make up a 1 liter solution.
(3) Dissolve using an ultrasonic vibrator.
(4) Pass through a 0.2microm membrane filter before use.

Sample :
Sodium azide, 30micro-L
1.
N3-, Azide ion 10ppm
2.
Cl-
Column : Shodex IC I-524A (4.6mmID*100mm)
Eluent : 2.5mM Phthalic acid(pH4.0 adjusted by Tris(hydroxymethyl)aminomethane)
Flow rate : 1.2mL/min
Detector : Shodex CD
Column temp. : 40deg-C