River Water (1) (I-524A)
Anions in river water were analyzed using IC
I-524A. In river water, small volume of H
2PO
4-, F
- and Br
- as well as Cl
-, NO
3- and SO
42- are present. This eluent can be used to separate H
2PO
4- and F
-.
(Preparation of eluent)
1) Put 0.332g of reagent grade phthalic acid, 0.223g of reagent grade Tris(hydroxymethyl)aminomethane
and 18.5g of boric acid into a 1L of messflask.
2) Dilute with purified water to make up a 1 liter solution.
3) Dissolve by stirring and warming.
4) Pass through a 0.2micron membrane filter before use.
This method is patented in Japan by Shimadzu Corporation.

Sample :
River water(
Water)
1.
H2PO4-
2.
F-
3.
Cl-
4.
Br-
5.
NO3-
6.
SO42-
Column : Shodex IC I-524A
Eluent : 2.0mM Phthalic acid + 1.84mM Tris(hydroxymethyl)aminomethane + 300mM Boric acid(pH4.0)
Flow rate : 1.2mL/min
Detector : Shodex CD
Column temp. : 40deg-C