Anion Standards (6) (I-524A)
Anion standards were analyzed using IC
I-524A. By adding
boric acid to the phthalic acid, phosphoric ion (H
2PO
4-)
and fluoride ion (F
-) can be separated.
(Preparation of eluent)
Put 0.332g of reagent grade phthalic acid, 0.223g of reagent grade tris(hydroxymethyl)aminomethane
and 18.5g of boric acid into a i liter messflask and dilute with pure water to
make up a 1 liter solution and dissolve it by stirring and warming.
This method is patented in Japan by Shimadzu Corporation.
Sample :
1.
H2PO4-
2.
F-
3.
Cl-
4.
NO2-
5.
Br-
6.
NO3-
7.
SO42-
Column : Shodex IC I-524A (4.6mmID*100mm)
Eluent : 2.0mM Phthalic acid + 1.84mM Tris(hydroxymethyl)aminomethane + 300mM Boric acid(pH4.0)
Flow rate : 1.2mL/min
Detector : Shodex CD
Column temp. : 40deg-C