Tap Water (4) (I-524A)
Based on the method described in the Japanese regulation for water supply, anions
were analyzed using IC
I-524A.
Please refer to '
Regulation for Water Supply'.
(Preparation of eluent)
1) Put 0.332g of reagent grade phthalic acid, 0.223g of reagent grade tris(hydroxymethyl)aminomethane
and 18.5g of boric acid into a 1 liter messflask.
2) Dilute with purified water to make up a 1 liter solution.
3) Dissolve by stirring and warming.
4) Pass through a 0.2micron membrane filter before use.
This method is patented in Japan by Shimadzu Corporation.

Sample :
Tap water (
Water)
1.
H2PO42-
2.
F-
3.
Cl-
4.
NO2-
5.
Br-
6.
NO3-
7.
SO42-
Column : Shodex IC I-524A (4.6mmID*100mm)
Eluent : 2.5mM Phthalic acid + 1.84mM Tris(hydroxymethyl)aminomethane + 300mM Boric acid(pH4.0)
Flow rate : 1.2mL/min
Detector : Shodex CD
Column temp. : 40deg-C