(Features)
| Model | ERC-3115alpha |
ERC-3215alpha |
ERC-3315alpha |
ERC-3415alpha |
|---|---|---|---|---|
| Solvent channels |
1 channel |
2 channels |
3 channels |
4 channels |
| Degassing capacity |
When ion-exchanged water saturated with air at 25-deg-C is put through at a flow rate of 3mL/min, no observed at the outlet of the apparatus. (Measured dissolved oxygen level at the outlet: 2ppm at flow rate of 3mL/min) |
|||
| Internal volume |
7mL/channel |
|||
| Max. flow rate | 20mL/min for each channel (Eluent: 25deg-C pure water) |
|||
| External output | An open connector signal is delivered to the external output signal terminal, When "PRES" and "LEAK" LEDs light. |
|||
| Dimensions | 71W x 310D x 136H (mm) |
|||
| Power source | AC 100 to 240 V, 50/60Hz |
|||
| Functional Displays |
Power On/Off display |
"POWER" LED lights, when the power is supplied. |
||
Status monitoring function |
"READY" LED lights, when the internal pressure in the vacuum chamber is below a predetermined limit. |
|||
Pressure monitoring function |
"PRES" LED lights, when the internal pressurein the vacuum chamber does not reach a predetermined level within a predetermined time. |
|||
Leak monitor |
"LEAK" LED lights, when the liquid leaks in the apparatus. |
|||
Self cleaning |
The vacuum line is cleaned by air suction. |
|||
Vacuum pump operation switching function |
NORM.: Controlled operation in normal run CONT.: Cntinuous operation in case og high degree of degassing degassing |
|||
| Weight |
ca 5.0kg |
ca 5.1kg |
ca 5.4kg |
ca 5.5kg |