Features of Multimode Columns (Peptide Analysis)
Although Asahipak
GS-HQ series is basically SEC columns,
they permit different types of separation and analysis from conventional ones
by use of a multimode which combines the hydrophobic interaction (RPC) and ionic
interaction (IEX: ion exclusion). The
GS-320 HQ column
performs well in separating hydrophillic peptides, particularly acidic and basic
peptides. As the columns can be used over a wide range of eluent pH because of
being polymer-based column, they can be washed with alkaline solution leading
to a longer column life, which is another feature of the column.
Sample :
1.
Val-Glu-Glu-Ala-Glu, 2.
Glu-Ala-Glu,
3.
Val-Glu-Ser-Glu
4.
Arg-Asp, 5.
Gly-His-Lys, 6.
Arg-Pro-Lys-Pro
Column : Shodex Asahipak GS-320 HQ (7.5mmID*300mm)
Eluent : 50mM Ammonium acetate buffer(pH6.7)
Flow rate : 0.5mL/min
Detector : UV(220nm)
Column temp. : 30deg-C