Features of RSpak DE Series (5) (Selection of HPLC System)

To realize the best performance of RSpak DE-413 column, we recommend to use the following HPLC system with the column.

Device Recommended specifications and/or setting Example
Pump Flow rate: max. 1mL/min.
Injector Low dead volume type
Sample loop: max. 50μL (5μL loop is recommended)

From injector
to column
0.1 to 0.13mmID, 100mm length piping
From column
to detector
0.1mmID piping
Detector Detector with a micro cell should be used.
Response should be set to "FAST".
Shodex RI-104
(small cell volume)



Sample : (A); 1.5μL, (B); 5μL
1. Methyl ethyl ketone, 2-Butanone 0.2%
2. Diethyl ketone, 3-Pentanone 0.2%
3. Di-n-propyl ketone, 4-Heptanone 0.4%
4. Di-n-butyl ketone, 5-Nonanone 0.6%

Column       : Shodex RSpak DE-413 (4.6mmID*150mm)
Eluent       : CH3CN/H2O=50/50
Flow rate    : 1.0mL/min
Detector     : UV(262nm)
Column temp. : 40°C