Glyphosate which is phosphorus-containing amino acid type herbicide was analyzed with IC/ICP-MS using IC SI-90 4C (a column for anion chromatography). Since SI-90 4C column has length as short as 100 mm, it can perform rapid analysis of glyphosate. In this experiment, after adding glyphosate into serum, it pretreated by centrifugal type extra filtration, and analyzed. as for a recovery rate, about 100% is obtained in the range of 0.1 to 100 ng/μ L -- detection marginal concentration was 0.1 ng/μL.
Shodex IC SI-90 4C is a special order.
Sample :
1. Glyphosate
Column : Shodex IC SI-90 4C (4.6 mm I.D. x 100 mm) Eluent : (A) H2O (B) 13 mM Na2CO3 + 20 mM NaHCO3 aq. Gradient: 0 to 3 min, A/B=90/10 3 to 15 min, A/B=90/10 to 70/30 Flow rate : 1.0 mL/min Suppresor : Dionex AMMS-ICS II Detector : ICP/MS, P at m/z=31 Column temp. : Room temp.