Trialkylammonium were separated using cation chromatography column, IC YK-421 . By using the mobile phase used below (high ACN concentration), hydrophobic interaction is suppressed. Thus ioninc-interaction dominates, this makes shorter chain alkylammonium to be retained longer than longer ones.


Sample : 100 mg/L each, 20 μL
1. Tetrahexylammonium
2. Tetrapentylammonium
3. Tetrabutylammonium
4. Tetrapropylammonium
5. Tetraethylammonium
6. Tetramethylammonium


Column       : Shodex IC YK-421 (4.6 mm I.D. x 125 mm)
Eluent       : 2.5 mM HNO3 aq./CH3CN=30/70
Flow rate    : 1.0 mL/min
Detector     : Non-suppressed coductivity
Column temp. : 40 °C