Cation standards and alkylamines can be analyzed simultaneously with IC YS-50.


Sample : 10 μL
1. Li+ 2 mg/L
2. Na+ 10 mg/L
3. NH4+ 10 mg/L
4. Methylamine 10 mg/L
5. K+ 20 mg/L
6. Trimethylamine, TMA 20 mg/L
7. Triethylamine, TEA 20 mg/L
8. Mg2+ 10 mg/L
9. Ca2+ 20 mg/L


Coefficient of determination (R2) (1 to 100 mg/L)
Methylamine 0.9997 Trimethylamine 0.9999 Triethylamine 0.9997
Column       : Shodex IC YS-50 (4.6 mm I.D. x 125 mm) 
Eluent       : 4 mM Methanesulfonic acid aq.
Flow rate    : 1.0 mL/min
Detector     : Non-suppressed conductivity
Column temp. : 40 °C