Acid selection depends on the sample type, instrument, and analysis conditions. Cation standards analyzed using IC YS-50 with 4 mM methanesulfonic acid, sulfuric acid, nitric acid, and phosphoric acid as the eluent are shouwn.


Sample : 10 μL
1. Li+ 2 mg/L
2. Na+ 10 mg/L
3. NH4+ 10 mg/L
4. K+ 20 mg/L
5. Mg2+ 10 mg/L
6. Ca2+ 20 mg/L


EluentConductivity
(μS/cm)
4 mM Sulfuric acid aq.
2906
4 mM Nitric acid aq.
1926
4 mM Methanesulfonic acid aq.
1733
4 mM Phosphoric acid aq.
1237
Column       : Shodex IC YS-50 (4.6 mm I.D. x 125 mm) 
Flow rate    : 1.0 mL/min
Detector     : Non-suppressed conductivity
Column temp. : 40 °C

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications

Applications (Related Information)