Analysis of Oxalic Acid Using Ion Exclusion Mode (KC-811)

Oxalic acid was analyzed using RSpak KC-811, a column for organic acid analysis. Oxalic acid, a dicarboxylic acid, is not retained on the column under standard eluent conditions (several mM perchloric acid solution) due to ion exclusion effects. It elutes in the same position as inorganic anions such as chloride ions, and thus prevents a quantitative analysis of oxalic acid. In such cases, increase of perchloric acid solution concentration suppresses the oxalic acid dissociation and let it retained in the column.  In this application, use of 50-mM aqueous perchloric acid solution enhanced the retention of oxalic acid and enabled its quantitative analysis without the influence of inorganic anions.

(Note)
When using a highly concentrated perchloric acid eluent as above, the post-column method using a pH indicator is not suitable. Detection by RI or UV (210 nm) is recommended.

Sample:

  1. 1.Anionic Compounds (Cl-)   10.Succinic acid
  2. 2.Oxalic acid                           11.Lactic acid
  3. 3.Citric acid                            12.Fumaric acid
  4. 4.Tartaric acid                        13.Acetic acid
  5. 5.Phosphoric acid                  14.Pyroglutamic acid
  6. 6.Malic acid
  7. 7.α-Ketoglutaric acid
  8. 8.Malonic acid
  9. 9.Pyruvic acid
  10.  
Column
Shodex RSpak KC-811 (8.0 mm I.D. x 300 mm) x 2
Eluent
50 mM HClO4 aq.
Flow rate
1.0 mL/min
Detector
UV (210 nm)
Column temp.
50 ℃