Xylooligosaccharides (2) (VN-50 4D)

Xylooligosaccharides were separated using HILICpak VN-50 4D, a polymer-based HILIC column, with evaporative light scattering detector (ELSD). Oligosaccharides up to 13-mer could be confirmed in this sample.

Sample: 10 μL
Xylooligosaccharides 5 mg/mL (in 50 % CH3CN)

  1. 1.Xylose                8.Xyloheptaose
  2. 2.Xylobiose           9.Xylooctaose
  3. 3.Xylotriose         10.Xylodecaose
  4. 4.Xylotetraose      11.Xyloundecaose
  5. 5.Xylopentaose    12.Xylododecaose
  6. 6.Xylohexose       13.Xylotridecaose
  7. 7.Xylohexaose
  8.  
Column
Shodex HILICpak VN-50 4D (4.6 mm I.D. x 150 mm)
Eluent
(A); 0.1 % NH3 aq. (B); CH3CN
Linear gradient; (B %) 80 % to 50 % (0 to 20 min)
Flow rate
1.0 mL/min
Detector
ELSD
Column temp.
40 ℃

Sample Name Index