Simultaneous Analysis of Amino Acids with LC/MS (VC-50 2D)

HILICpak VC-50 2D is a polymer-based amino HILIC column that was used to analyze 20 amino acids. This method does not require the use of derivatization or ion-pair reagents. The result showed basic amino acids were separated well and tend to retain longer by the carboxyl functional group modified on VC-50 2D column packing material. When analyzing acidic amino acids, we recommend HILICpak VG-50 2D.
Refer to: Simultaneous Analysis of Saccharides, Organic Acids and Amino Acids with LC/MS (VG-50 2D)

Sample: 5 μL
Amino acid standards 0.1 μg/mL each (in H2O)

Scroll
Aspartic acid, Asp Glutamic acid, Glu Cysteine, Cys Threonine, Thr
Serine, Ser Proline, Pro Tyrosine, Tyr Asparagine, Asn
Methionine, Met Glutamine, Gln Lysine, Lys Alanine, Ala
Glycine, Gly Phenylalanine, Phe Valine, Val Leucine, Leu
Isoleucine, Ile Tryptophan, Trp Histidine, His Arginine, Arg
Column
Shodex HILICpak VC-50 2D (2.0 mm I.D. x 150 mm)
Eluent
(Low pressure linear gradient)
H2O/100 mM HCOOH aq./CH3CN=30/20/50 to 0/90/10 (0 to 5 min), 0/90/10 (5 to 30 min)
Flow rate
0.2 mL/min
Detector
ESI-MS (SIM)
Column temp.
30 ℃