Analysis of Peptide using Multimode Column

Although Asahipak GS-HQ series is basically SEC columns, they permit different types of separation and analysis from conventional ones by use of a multimode which combines the hydrophobic interaction (RPC) and ionic interaction (IEX: ion exclusion). The GS-320 HQ column performs well in separating hydrophillic peptides, particularly acidic and basic peptides. As the columns can be used over a wide range of eluent pH because of being polymer-based column, they can be washed with alkaline solution leading to a longer column life, which is another feature of the column.

Sample: 20 µL

  1. 1.Glu-Ala-Glu 0.025 %
  2. 2.Arg-Asp 0.05 %
  3. 3.Gly-His-Lys 0.025 %
  4. 4.Arg-Pro-Lys-Pro 0.025 %
Scroll
Peptides MW pI Σf
Glu-Ala-Glu 347   3.12 0.39
Arg-Asp 289   6.75 0.68
Gly-His-Lys 340   9.95 0.29
Arg-Pro-Lys-Pro 497 11.44 3.24
PI: Isocratic point
Σf: Hydrophobic parameter
Column
Shodex Asahipak GS-320 HQ (7.5 mm I.D. x 300 mm)
Eluent
50 mM Ammonium acetate buffer (pH6.7)
Flow rate
0.5 mL/min
Detector
UV (220 nm)
Column temp.
30 ℃

Sample Name Index

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Applications