Ligand Exchange Chromatography Columns (Ligand Exchange and HILIC)

Print
DC-613
SZ5532
SC1211
· Separates elements by combination of ligand exchange and HILIC modes
· DC-613 can analyze sugars without removing sodium salts in the sample
· SZ5532 is recommended for the separation of disaccharides or trisaccharides
· SC1211 is suitable for separating sugar alcohols
· DC-613 fulfills USP L22 and L58 requirements
· SZ5532 fulfills USP L22 requirements
· SC1211 fulfills USP L19 and L22 requirements

(Standard columns)
Product codeProduct namePlate number
(TPN/column)
Functional group
(Counter ion)
Particle size
(μm)
Size (mm)
I.D. x Length
F7001003
F6700170
RSpak DC-613
RSpak DC-G 4A
≥ 5,500
(guard column)
Sulfo (Na+)
Sulfo (Na+)
6
10
6.0 x 150
4.6 x 10
F7001300
F6700110
SUGAR SZ5532
SUGAR SZ-G
≥ 5,500
(guard column)
Sulfo (Zn2+)
Sulfo (Zn2+)
6
6
6.0 x 150
4.6 x 10
F7001400
F6700120
SUGAR SC1211
SUGAR SC1211G 4A
≥ 5,500
(guard column)
Sulfo (Ca2+)
Sulfo (Ca2+)
6
10
6.0 x 250
4.6 x 10
Product nameMax. usable
pressure (MPa)
Recommended
flow rate
(mL/min)
Max. usable
flow rate
(mL/min)
Usable
temp. (°C)
Shipping solvent
DC-613
DC-G 4A

5.0
-

0.5 to 1.0
-

3.0
-

40 to 85
40 to 85

CH3CN/H2O=70/30
CH3CN/H2O=70/30

SZ5532
SZ-G
4.0
-
0.5 to 1.0
-
2.0
-
≤ 80
≤ 80
CH3CN/H2O=70/30
CH3CN/H2O=70/30
SC1211
SC1211G 4A
4.0
-
0.5 to 1.0
-
1.0
-
≤ 95
≤ 95
CH3CN/H2O=25/75
CH3CN/H2O=25/75
Eluent condition for DC-613: Please refer to Eluent conditions for RSpak D sereies.