Polymer-based Columns for Hydrophilic Interaction Chromatography (HILIC) (HILICpak)

Scroll
VG-50
  • Suitable for saccharide analysis using HILIC mode
  • Recovers reducing saccharides with high ratio
  • Polymer-based packing material provides excellent chemical stability and minimum deterioration over an extended time period
  • Easily regenerated by washing in an alkaline solution
  • Appropriate for evaporative light scattering detector, corona charged aerosol detector, and LC/MS
VT-50 2D
  • Suitable for anionic substances (especially phosphate compounds) analysis using HILIC mode
  • Use of some eluents add ion exchange mode
  • Polymer-based packing material provides excellent chemical stability and minimum deterioration over an extended time period
  • Suitable for LC/MS analysis
VC-50 2D
  • Modified carboxyl group is suitable for cationic substance analysis including amines
  • The dominant separation mode is RP or IEX rather than HILIC mode
VN-50
  • The modified diol groups on the packing material create the HILIC mode
  • Suitable for oligonucleotide and oligosaccharide separation which is not possible by SEC or conventional HILIC columns
  • VN-50 1D adopts a double structure housing with an outer surface made of SUS and an inner surface [wetted part] made of PEEK. Covers the weak point of physical strength of PEEK housing and can be used under high pressure conditions.
  • VN-50 1D has an inner diameter of 1.0 mm, which makes it possible to suppress the flow rate to 0.1 mL/min or less, so it suitable for high sensitivity analysis by LC/MS.
HILICpak VG-50(Standard columns)   [Housing material: SUS]Scroll
Product code Product name Plate number
(TPN/column)
Functional group Particle size (μm) Column size (mm)
I.D. x Length
F7630200 HILICpak VG-50 4D ≥ 5,500 Amino 5 4.6 x 150
F7630100 HILICpak VG-50 4E ≥ 7,500 Amino 5 4.6 x 250
F6711100 HILICpak VG-50G 4A (guard column) Amino 5 4.6 x   10
Scroll
Product name Maximum pressure
(MPa/Column)
Flow rate (mL/min) Temperature range
(℃)
pH range Shipping solvent
Recommended Maximum
HILICpak VG-50 4D 10 0.5 - 1.0 1.5 4 - 60 2 - 13 H2O/CH3CN=20/80
HILICpak VG-50 4E 15 0.5 - 1.0 1.5 4 - 60 2 - 13 H2O/CH3CN=20/80
HILICpak VG-50G 4A - - - 4 - 60 2 - 13 H2O/CH3CN=20/80
(Semi-micro columns)   [Housing material: PEEK] Scroll
Product code Product name Plate number
(TPN/column)
Functional group Particle size (μm) Column size (mm)
I.D. x Length
F7630300 HILICpak VG-50 2D ≥ 3,500 Amino 5 2.0 x 150
F6711200 HILICpak VG-50G 2A (guard column) Amino 5 2.0 x   10
Scroll
Product name Maximum pressure
(MPa/Column)
Flow rate (mL/min) Temperature range
(℃)
pH range Shipping solvent
Recommended Maximum
HILICpak VG-50 2D 15 0.1 - 0.2 0.45 4 - 60 2 - 13 H2O/CH3CN=15/85
HILICpak VG-50G 2A - - - 4 - 60 2 - 13 H2O/CH3CN=15/85
Eluents:
Aqueous solutions of Acetonitrile (or Methanol) at any ratio, is used as eluent.
Any salts soluble in the solution above are available.
HILICpak VT-50(Semi-micro columns)   [Housing material: PEEK]Scroll
Product code Product name Plate number
(TPN/column)
Functional group Particle size (μm) Column size (mm)
I.D. x Length
F7630400 HILICpak VT-50 2D ≥ 4,500 Quaternary ammonium 5 2.0 x 150
F6711300 HILICpak VT-50G 2A (guard column) Quaternary ammonium 5 2.0 x   10
Scroll
Product name Maximum pressure
(MPa/Column)
Flow rate (mL/min) Temperature
range (℃)
pH range Shipping solvent
Recommended Maximum
HILICpak VT-50 2D 10 0.2 - 0.3 0.5 4 - 60 2 - 13

25 mM HCOONH4 aq./CH3CN=15/85

HILICpak VT-50G 2A - - - 4 - 60 2 - 13

25 mM HCOONH4 aq./CH3CN=15/85

Eluents:
Aqueous solutions of Acetonitrile (or Methanol) at any ratio, is used as eluent.
Any salts soluble in the solution above are available.
HILICpak VC-50(Semi-micro columns)   [Housing material: PEEK] Scroll
Product code Product name Plate number
(TPN/column)
Functional group Particle size (μm) Column size (mm)
I.D. x Length
F7630700 HILICpak VC-50 2D ≥ 3,500 Carboxyl 5 2.0 x 150
F6711600 HILICpak VC-50G 2A (guard column) Carboxyl 5 2.0 x   10
Scroll
Product name Maximum pressure
(MPa/Column)
Flow rate (mL/min) Temperature range
(℃)
pH range Shipping solvent
Recommended Maximum
HILICpak VC-50 2D 10 0.1 - 0.3 0.5 4 - 60 2 - 12 H2O
HILICpak VC-50G 2A - - - 4 - 60 2 - 12 H2O
Eluents:
Aqueous solutions of Acetonitrile at any ratio, is used as eluent.*
Any salts soluble in the solution above are available.
*Methanol is not available as the eluent. Other organic solvents including Methanol are not guaranteed as the eluent.
HILICpak VN-50(Standard columns)   [Housing material: PEEK]Scroll
Product code Product name Plate number
(TPN/column)
Functional group Particle size (μm) Column size (mm)
I.D. x Length
F7630500 HILICpak VN-50 4D ≥ 10,000 Diol 5 4.6 x 150
F6711400 HILICpak VN-50G 4A (guard column) Diol 5 4.6 x   10
Scroll
Product name Maximum pressure
(MPa/Column)
Flow rate (mL/min) Temperature range
(℃)
pH range Shipping solvent
Recommended Maximum
HILICpak VN-50 4D 10 0.3 - 1.0 1.5 4 - 60 2 - 13 H2O/CH3CN=25/75
HILICpak VN-50G 4A - - - 4 - 60 2 - 13 H2O/CH3CN=25/75
(Semi-micro columns)   [Housing material: Double structure with outer surface SUS and inner surface (wetted part) PEEK]  *VN-50 1D is made to order. Scroll
Product code Product name Plate number
(TPN/column)
Functional group Particle size (μm) Column size (mm)
I.D. x Length
F7630800 HILICpak VN-50 1DNEW ≥ 2,800 Diol 5 1.0 x 150
Scroll
Product name Maximum pressure
(MPa/Column)
Flow rate (mL/min) Temperature range
(℃)
pH range Shipping solvent
Recommended Maximum
HILICpak VN-50 1D 10 0.01 - 0.05 0.1 4 - 60 2 - 13 H2O/CH3CN=25/75
[Housing material: PEEK] Scroll
Product code Product name Plate number
(TPN/column)
Functional group Particle size (μm) Column size (mm)
I.D. x Length
F7630600 HILICpak VN-50 2D ≥ 3,500 Diol 5 2.0 x 150
F6711500 HILICpak VN-50G 2A (guard column) Diol 5 2.0 x   10
Scroll
Product name Maximum pressure
(MPa/Column)
Flow rate (mL/min) Temperature range
(℃)
pH range Shipping solvent
Recommended Maximum
HILICpak VN-50 2D 10 0.1 - 0.3 0.5 4 - 60 2 - 13 H2O/CH3CN=25/75
HILICpak VN-50G 2A - - - 4 - 60 2 - 13 H2O/CH3CN=25/75
(Preparative columns) [Housing material: SUS (VN-50 10E), PEEK (VN-50G 4A)]  *Preparative columns are made to order. Scroll
Product code Product name Plate number
(TPN/column)
Particle size (μm) Column size (mm)
I.D. x Length
Shipping solvent
F6830100 HILICpak VN-50 10E ≥ 11,000 5 10.0 x 250 H2O/CH3CN=25/75
F6711400 HILICpak VN-50G 4A (guard column) 5  4.6 x   10 H2O/CH3CN=25/75
Eluents:
Aqueous solutions of Acetonitrile (or Methanol) at any ratio, is used as eluent.
Any salts soluble in the solution above are available.

Operation Manual / Certificate of Analysis

Operation Manuals and Certificate of Analysis / Inspection Certificate for the following products can be downloaded here.

Product Name Index

Related Products

Product Information (Related Information)