聚合物反相柱ODP2 HP-4D通碱液超过500小时后,各样品的保留时间和理论塔板数基本不改变,显示了优异的耐碱性。然而,ODS (A)通碱液24小时后,各样品的保留时间逐渐变快,柱效降低,表明色谱柱劣化。
Test condition of alkali durability: Eluent : 10 mM Phosphate buffer (pH12)/CH3CN=45/55 Flow rate : 0.6 mL/min Column temp. : 30 ℃
Sample : 5 μL
1. Pyridine 200 mg/L
2. Phenol 430 mg/L
* TPN before flowing alkali eluent is set as 100% in this test.
Column : Shodex ODP2 HP-4D (4.6 mm I.D. x 150 mm) ODS(A) from other manufacturer (4.6 mm I.D. x 150 mm) Eluent : H2O/CH3OH=70/30 Flow rate : 1.0 mL/min Detector : UV (254 nm) Column temp. : 40 ℃